HMX Series pressure transmitters with HART communication protocol for use in high temperature environments with explosive atmosphere presence
Exposed Capillary
• Pressure ranges from: 0-17 a 0-2000 bar/0-250 a 0-30000 psi
• Extensimetric measurement principle with Wheatstone bridge
• Calibration signal 80% FSO internally generated
• Completely interchangeable with all existing products
• Protection level: IP65 (6-pin connector)
• 1/2-20UNF, M18x1.5 standard threads; other types available
on request
• Standard diaphragm is 15-5 PH stainless steel with GTP coating
• 17-7 PH corrugated stainless steel diaphragm with GTP coating
for ranges below 100 bar-1500 psi
• Other diaphragm types available on request
• Accuracy: < ±0.25% FSO (H); < ±0.5% FSO (M)